VALIDATED LC-MS/MS METHOD FOR THE DETERMINATION OF THE MUSCARINIC RECEPTOR ANTAGONIST (MRA) SOLIFENACIN FROM HUMAN PLASMA

Authors

  • Róbert TŐTŐS Biocatalysis and Biotransformations Research Group, Faculty of Chemistry and Chemical Engineering, Babeş-Bolyai University, Cluj-Napoca, Romania. Email: robert.totos@ubbcluj.ro. https://orcid.org/0000-0003-1743-3905
  • József BALÁZSI PharmacoKinetics SRL, Corunca, Mures county, Romania. Corresponding author: robert.totos@ubbcluj.ro.

DOI:

https://doi.org/10.24193/subbchem.2019.2.24

Keywords:

solifenacin, muscarinic antagonist, method validation, bioequivalence trial, LC-MS/MS

Abstract

The purpose of this study was the development and validation of an LC-MS/MS method, for the determination of solifenacin from human plasma. The sample workup involved a simple protein precipitation procedure. A core/shell type analytical column (50×2,1 mm, 2.6 Å) was used with C18 stationary phase. The mobile phase consisting of 65% acetonitril and 35% water provided good peak shape, accuracy and precision. The mass spectrometer was operated in positive electrospray ionization mode for analyte and internal standard. The following parameters were evaluated for validation purpose: Selectivity, sensitivity, matrix effect, anticoagulant effect, linearity, precision and accuracy, recovery, short and long term analyte/IS stability in solvent/matrix and carryover. The validated calibration range was 0.71-71.28 ng/ml. The correlation coefficient R2 was at least 0.99 in all validation batches. The validated method has been successfully used for the evaluation of bioequivalence of generic solifenacin 10 mg formulations.

References

https://www.rxlist.com/vesicare-drug.htm: VESIcare (solifenacin succinate) Tablets; Viewed: 02.11.2018).

O. Doroshyenko, U. Fuhr, Clin. Phramacokinet., 2009, 48, 281.

L. Singh, S. Nanda, Pharmaceutical Methods, 2011, 2, 21.

H.N. Mistri, A.G. Jangid, A. Pudage, D.M. Rathod, P.S. Shrivastav, Journal of Chromatography B, 2008, 876, 236.

J. Macek, P. Ptáček, J. Klíma, Jorunal of Chromatography B, 2010, 878, 3327.

T. Uchida, W.J. Krauwinkel, H. Mulder, R.A. Smulders, Br. J. Clin. Pharmacol., 2004, 58, 4.

Yanagihara T, Aoki T, Soeishi Y, Iwatsubo T, Kamimura H., Journal of Chromatography B., 2007, 859, 241.

EMEA/CHMP/EWP/192217/2009 Rev. 1 Corr. 2** Guideline on validation of bioanalytical method, 21 July 2011 (Updated 03/06/2015).

Downloads

Published

2019-06-03

How to Cite

TŐTŐS, R. ., & BALÁZSI, J. . (2019). VALIDATED LC-MS/MS METHOD FOR THE DETERMINATION OF THE MUSCARINIC RECEPTOR ANTAGONIST (MRA) SOLIFENACIN FROM HUMAN PLASMA. Studia Universitatis Babeș-Bolyai Chemia, 64(2), 289–296. https://doi.org/10.24193/subbchem.2019.2.24

Issue

Section

Articles

Similar Articles

<< < 8 9 10 11 12 13 14 15 16 17 > >> 

You may also start an advanced similarity search for this article.